Contact an Account ManagerStandard SPE Procedure for CHROMABOND® HILIC
Summary
Type: SPE
Application #: 305580
Catalog#: 730593 - SPE Cartridges, Chromabond, HILIC, 3ml, PP, 500mg 50/pk
Detection:
Conditions: Column conditioning: 1 mL water
(Do not let run the column dry!)
Equilibration: 6 mL acetonitrile or the organic solvent, diluted the sample
Sample aspiration: prepared sample is passed dropwise through the column
Washing: if necessary, 0.5–2 mL acetonitrile or the organic solvent, diluted the sample
Elution: 1–2 mL water (dependent on analyt)
Evaporation and reconstitution (if necessary)
Authors: SPE department
Source: Macherey-Nagel, Germany 2012